H1 OMX Startup and Shutdown manual
From BioDIP
(Difference between revisions)
(details of user power up) |
|||
Line 1: | Line 1: | ||
Start up and Shut down manual for the OMX system MPI-LMF H1 | Start up and Shut down manual for the OMX system MPI-LMF H1 | ||
− | Please ask LMF for training. | + | ==General notes== |
+ | Please ask LMF for training. | ||
+ | See section 3.3 of OMX manual at the system | ||
+ | *The manual explains the full power on and off procedures, but might be out of date. | ||
+ | *On this Wiki page is written the normal user power on/off procedures | ||
+ | ** Normal users do not fully power off and on the system | ||
+ | ** Certain components are usually left on for stability reasons: galvos, coolers, camera an OMXIC computers etc. | ||
+ | ==Normal user System Turn ON== | ||
+ | #Open the doors of the laser unit to see what is already on. | ||
+ | #Make sure all computers are on and booted up, and logged in. | ||
+ | ## 4x camera computers plus the OMXIC computer in the laser unit | ||
+ | ## Both the OMX master and the Linux reconstruction workstation computers. | ||
+ | #Power ON the OMX hardware control and laser systems in the laser unit: | ||
+ | ##Nano motion Chassis - right side | ||
+ | ##OMX Instrument Controller Chassis - right side | ||
+ | ### make sure the despeckle unit has power (green light next to the switch number 1 - it should make a noise and vibrate the laser unit) | ||
+ | ##Peizo z controller (bottom left of the unit - small power switch on front of controller - bottom right) | ||
+ | #Power on the 2 laser control modules on the bottom left | ||
+ | ## Main power of both boxes, then also turn on each laser power under the label for each laser line. | ||
+ | ## Turn the interlock key to position 1 - so lader light is allowed out of the system | ||
+ | # The light on the laser unit door and the OMX enclosure should be green (means powered on but not active yet.) | ||
+ | # Make sure you are logged into the OMX master windows computer (the left one) | ||
+ | # Start the OMX software | ||
+ | #restart the OMX hardware | ||
+ | ##menu item: Hardware - Hardware | ||
+ | ## Click the restart hardware" button - wait for a couple of min. | ||
+ | ## when its done there should be no red text at the hardware and camera status displays. | ||
+ | ### and the camera icon should not be greyes out - and you should be able to snap and image from one or all the cameras. | ||
+ | |||
+ | ==Normal user System Turn OFF== | ||
+ | |||
+ | == Trouble shooting== | ||
+ | Are the 4 galvo controllers powered on? | ||
+ | |||
+ | ==Full System Power ON== | ||
+ | |||
+ | ==Full System Power OFF== | ||
+ | |||
+ | |||
+ | ==Recent changes== | ||
New as of Blaze upgrade: | New as of Blaze upgrade: | ||
* 2x 2 galvo control boxes. 2 on OMX enclosure, 2 on laser cabinet. | * 2x 2 galvo control boxes. 2 on OMX enclosure, 2 on laser cabinet. |
Revision as of 11:58, 11 January 2012
Start up and Shut down manual for the OMX system MPI-LMF H1
Contents |
General notes
Please ask LMF for training. See section 3.3 of OMX manual at the system
- The manual explains the full power on and off procedures, but might be out of date.
- On this Wiki page is written the normal user power on/off procedures
- Normal users do not fully power off and on the system
- Certain components are usually left on for stability reasons: galvos, coolers, camera an OMXIC computers etc.
Normal user System Turn ON
- Open the doors of the laser unit to see what is already on.
- Make sure all computers are on and booted up, and logged in.
- 4x camera computers plus the OMXIC computer in the laser unit
- Both the OMX master and the Linux reconstruction workstation computers.
- Power ON the OMX hardware control and laser systems in the laser unit:
- Nano motion Chassis - right side
- OMX Instrument Controller Chassis - right side
- make sure the despeckle unit has power (green light next to the switch number 1 - it should make a noise and vibrate the laser unit)
- Peizo z controller (bottom left of the unit - small power switch on front of controller - bottom right)
- Power on the 2 laser control modules on the bottom left
- Main power of both boxes, then also turn on each laser power under the label for each laser line.
- Turn the interlock key to position 1 - so lader light is allowed out of the system
- The light on the laser unit door and the OMX enclosure should be green (means powered on but not active yet.)
- Make sure you are logged into the OMX master windows computer (the left one)
- Start the OMX software
- restart the OMX hardware
- menu item: Hardware - Hardware
- Click the restart hardware" button - wait for a couple of min.
- when its done there should be no red text at the hardware and camera status displays.
- and the camera icon should not be greyes out - and you should be able to snap and image from one or all the cameras.
Normal user System Turn OFF
Trouble shooting
Are the 4 galvo controllers powered on?
Full System Power ON
Full System Power OFF
Recent changes
New as of Blaze upgrade:
- 2x 2 galvo control boxes. 2 on OMX enclosure, 2 on laser cabinet.