How to start the Zeiss LSM UV

Control Panel

Remote control.png

  1. Switch on Remote Control

Software Initiation

Icon.png Switchboard.png

  1. Start the software
  2. Make sure "Scan New Images" is clicked
  3. Then press "Start Expert Mode"

Basic Acquisition Setup

Software menu.png

  1. This is the main software window. From here you can access all needed control windows (bottom row buttons)

Laser control.png Microscope control.png Config control.png Scan-mode.png Scan control.png

  1. Laser Control
    Switch on the lasers you need.
    Remember: for Argon laser use tube current of around 6A
  2. Microscope Control
    Here you can make changes to the microscope stand itself.
    Note: most of these functions you can also directly access via button on the microscope stand.
  3. Conficuration Control
    Here you can set up your beam path configuration.
    Use "Single Track" for simultaneous, "Multi Track" for sequential imaging
    To set up everything properly, you need to know the excitation and emission spectra of each of your dyes. Please use the provided fluorescence database links
  4. Scan Control
    • Mode
      Adjust your scan settings, such as pixel size, scan speed, averaging, zoom,...
    • Channels
      Adjust pinhole size, detector gain, amplifier offset, and laser power for each channel separately

Observation Modes

Switch.png

Use the buttons on the right side of the main software window to switch between visual (VIS; observation by eye) and Laser Scanning Mode (LSM; observation over displays)