Camera computers - RDP | Camera computers - RDP | ||
ntservice - process that talks to the omx master PC | ntservice - process that talks to the omx master PC | ||
+ | |||
+ | Camera Offset setting is in File - Settings | ||
+ | |||
+ | Chip pixel correction - is in Hardware Test : PCO Camera Bias (On/Off) Careate Bias image - takes a few min to capture and average dark images for each frame size. Default is 10 images, and all frame sizes. |
Revision as of 12:52, 16 December 2011
Notes for admin and advanced stuff.
Camera computers - RDP ntservice - process that talks to the omx master PC
Camera Offset setting is in File - Settings
Chip pixel correction - is in Hardware Test : PCO Camera Bias (On/Off) Careate Bias image - takes a few min to capture and average dark images for each frame size. Default is 10 images, and all frame sizes.