L5 - Olympus ScanR+CellR

From BioDIP
Revision as of 17:33, 16 October 2008 by Old user (Talk | contribs)

Jump to: navigation, search
Suitable for long-term observation of living cells.
MPI24 .jpg


Directions

[[image:|150px|text-bottom|right]]

[[image:|100px]] MPI-CBG LMF (MPI-CBG, U50)

Booking

https://python-srv1.mpi-cbg.de/lmf-ipf/cgi-bin/index.py

Details

microscope , Inverted stand, motorized XY stage, motorized Z-drive, fluorescence, transmitted light
objectives
illumination
detection Hamamatsu Camera ORCA ER, progressive scan interline CCD, 1344 x 1024 pixels, 6.45 um pixel size, 14.75 MHz digitization, 8 e- rms readout noise, 8 fps @ full frame, 18.000 e- full well capacity, 2250:1 dynamic range, peltier cooling (-20 degrees C @ 20 degrees C ambient), 12 bit ADC, 0.1 electron/pixel/sec dark current, up to 10x analog gain, offset functions
reflectors MT20 Excitation Unit
- EXCITATION FILTERS
403/ 492 / 572 / 387/ 485 / 560/ 650
- DICHROICS AND EMISSION FILTERS
features -
software
incubation cage incubator with temperature and CO2 control capabilities
links
  • [ manual]
  • [- laser power values]
  • [- beam path]
  • [ advanced system check]
inv.nr. 405679



[[Category:{{{type}}}]]