LSM 510, inverse, BIOTEC

From BioDIP
Revision as of 15:34, 24 November 2010 by Elleng (Talk | contribs)

Jump to: navigation, search
Suitable for - confocal scanning microscopy.
Bioz01.jpg


Directions

[[image:|150px|text-bottom|right]]

[[image:|100px]] Imaging@Biotec"Imaging@Biotec" is not in the list of possible values (DZNE - Imaging Platform, EM- and Histo-Facility CRTD / BIOTEC, Image Processing@BIOTEC, Imaging@BIOTEC/CRTD, MPI-CBG LMF, MPI-CBG LMF - Quantitative Image Analysis Unit, MPI-CBG TDS, MTZ Imaging, MPI-CBG EM, BioDIP) for this property. ("Biotec" is not in the list of possible values (BIOTEC, CRTD, DZNE, MTZ, MPI-CBG) for this property., 221)

Booking

https://techpool.biotec.tu-dresden.de/schedule/schedule.php?scheduleid=sc142df4c59eef5c

Details

microscope Zeiss - Axiovert 200M, Inverted stand, inverted stand, motorized XY stage, motorized Z-drive, fluorescence, transmitted light with manual DIC
objectives
Zeiss EC Plan-Neofluar 10x 0.3
Zeiss EC Plan-Neofluar 20x 0.5
Zeiss C-Apochromat 40x 1.2 W
Zeiss C-Apochromat 63x 1.2 W
Zeiss Plan-Apochromat 63x 1.4 Oil
Zeiss alpha Plan-Fluar 100x 1.45 Oil
illumination
Transmitted Light (Halogen)
Fluorescence (HBO)
Laser Diode 405 nm
Laser Argon Multiline 458, 477, 488, 514 nm
Laser HeNe 543 nm
Laser HeNe 633 nm
detection point scanner, 2 confocal PMTs, 1 META detector, T-PMT, dichromatic mirrors for excitation/emission splitting, emission filters for detection range selection, pinhole for each detector
reflectors empty
DAPI : EX 365/12 ; BS 395 ; EM LP 397
FITC : EX 470/40 ; BS 510 ; EM LP 515
TRITC : EX 546/12 ; BS 580 ; EM LP 590 LP
empty
empty
features simultaneous imaging, sequential imaging, line scan, point scan, Z stacks, time-series, advanced time-series, averaging, summarizing, scan zoom and rotation, FRAP
software Zeiss AIM
incubation Stage incubator (T)
links
  • [- beam path]
  • [ advanced system check]
inv.nr. -