SD1 - Andor Spinning Disc

From BioDIP
(Difference between revisions)
Jump to: navigation, search
(added filters in pos 7 and 8 in filterwheel)
Line 22: Line 22:
 
|ill4=Laser DPSS 561 nm
 
|ill4=Laser DPSS 561 nm
 
|detection=Spinning disc scan head Yokogawa CSU10: dichromatic mirror: [http://www.semrock.com/Catalog/Detail.aspx?FilterPartID=855&CategoryID=82 T-405/488/568/647]<br> physical pinhole radius: 24um, physical pinhole spacing: 240um; back-projected pinhole radius: 0.15um, back-projected pinhole spacing: 1.5um (with 100x objective, 1.6x optovar)<br>Andor iXon EM+ DU-897 BV back illuminated EMCCD; pixel size of EMCCD chip: 16um<br>image pixel size (measured): with 100x objective: 0.178um; 100x obj. and 1.6x optovar: 0.111um; 60x obj.: 0.288um; 60x obj. and 1.6x opt.: 0.181um
 
|detection=Spinning disc scan head Yokogawa CSU10: dichromatic mirror: [http://www.semrock.com/Catalog/Detail.aspx?FilterPartID=855&CategoryID=82 T-405/488/568/647]<br> physical pinhole radius: 24um, physical pinhole spacing: 240um; back-projected pinhole radius: 0.15um, back-projected pinhole spacing: 1.5um (with 100x objective, 1.6x optovar)<br>Andor iXon EM+ DU-897 BV back illuminated EMCCD; pixel size of EMCCD chip: 16um<br>image pixel size (measured): with 100x objective: 0.178um; 100x obj. and 1.6x optovar: 0.111um; 60x obj.: 0.288um; 60x obj. and 1.6x opt.: 0.181um
|reflectors=Optical Insight DualView image splitter: <br>HC 520/35; BS 565; HC 628/40<br>Sutter filter wheel: <br>position 0: 624/40,  1: open,  2: 525/30,  3: razor edge 568LP,  4: dual emitter: 512/23+630/91,  5: razor edge 488LP,  6: old CSU dual emitter:  525/30 + 650/?,  7: blocked,  8: blocked,  9: 716/40<br> microscope filter turret:<br> position 1: GFP cube,  2: RFP cube,  3: dual GFP + RFP cube,  4: DAPI cube,  5: empty,  6: DIC analyzer
+
|reflectors=Optical Insight DualView image splitter: <br>HC 520/35; BS 565; HC 628/40<br>Sutter filter wheel: <br>position 0: 624/40,  1: open,  2: 525/30,  3: razor edge 568LP,  4: dual emitter: 512/23+630/91,  5: razor edge 488LP,  6: old CSU dual emitter:  525/30 + 650/?,  7: 605/70,  8: 445/40,  9: 716/40<br> microscope filter turret:<br> position 1: GFP cube,  2: RFP cube,  3: dual GFP + RFP cube,  4: DAPI cube,  5: empty,  6: DIC analyzer
 
|features=-
 
|features=-
 
|software=iQ 1.9.1
 
|software=iQ 1.9.1

Revision as of 13:51, 30 September 2010

Suitable for confocal imaging of EGFP and mCherry with high temporal and spatial resolution..
MPI19.jpg


Directions

[[image:|150px|text-bottom|right]]

[[image:|100px]] MPI-CBG LMF (MPI-CBG, 038B)

Booking

https://python-srv1.mpi-cbg.de/lmf-ipf/cgi-bin/index.py

Details

microscope Olympus - IX71, Inverted stand, manual, 1.6x OptoVar, fast piezo z-drive for objective
objectives
Olympus UPlanSApo 10x 0.4
Olympus UPlanSApo 60x 1.20 W
Olympus UPlanSApo 100x 1.4 Oil
illumination
Transmitted Light (Halogen)
Fluorescence (HBO)
Laser DPSS 488 nm
Laser DPSS 561 nm
detection [[detection::Spinning disc scan head Yokogawa CSU10: dichromatic mirror: T-405/488/568/647
physical pinhole radius: 24um, physical pinhole spacing: 240um; back-projected pinhole radius: 0.15um, back-projected pinhole spacing: 1.5um (with 100x objective, 1.6x optovar)
Andor iXon EM+ DU-897 BV back illuminated EMCCD; pixel size of EMCCD chip: 16um
image pixel size (measured): with 100x objective: 0.178um; 100x obj. and 1.6x optovar: 0.111um; 60x obj.: 0.288um; 60x obj. and 1.6x opt.: 0.181um]]
reflectors Optical Insight DualView image splitter:
HC 520/35; BS 565; HC 628/40
Sutter filter wheel:
position 0: 624/40, 1: open, 2: 525/30, 3: razor edge 568LP, 4: dual emitter: 512/23+630/91, 5: razor edge 488LP, 6: old CSU dual emitter: 525/30 + 650/?, 7: 605/70, 8: 445/40, 9: 716/40
microscope filter turret:
position 1: GFP cube, 2: RFP cube, 3: dual GFP + RFP cube, 4: DAPI cube, 5: empty, 6: DIC analyzer
features -
software iQ 1.9.1
incubation Stage incubator (T) + objective heater
links
  • [- beam path]
  • [ advanced system check]
inv.nr. 0



Personal tools