Line 26: Line 26:
 
|detection=Spinning disc scan head Yokogawa CSU-X1 (10.000rpm):<br>dichromatic mirrors:<br>[http://www.semrock.com/FilterDetails.aspx?id=Di01-T405/488/568/647-13x15x0.5 Quad band T-405/488/568/647]<br> [http://www.semrock.com/FilterDetails.aspx?id=Di01-T405/488/561-13x15x0.5 Triple band T-405/488/561] <br>physical pinhole radius: 24um, physical pinhole spacing: 240um; back-projected pinhole radius: 0.15um, back-projected pinhole spacing: 1.5um (with 100x objective, 1.6x optovar)<br>Andor iXon EM+ DU-897 BV back illuminated EMCCD; pixel size of EMCCD chip: 16um<br>image pixel size (measured again after major scope repair on March 7th 2011): with 100x objective: 0.175um (was 0.178um before 20110307); 100x obj. and 1.6x optovar: 0.109um (was 0.111um before 20110307); 60x obj.: 0.289um (was 0.288um before); 60x obj. and 1.6x opt.: 0.180um (was 0.181um before 20110307); 10x obj. : 1.747um, 10x obj. and 1.6x opt.: 1.089um<br>camera projecting lens has about 0.92x (was 0.9x before 2011_03_07) magnification calculated from the measured pixel sizes
 
|detection=Spinning disc scan head Yokogawa CSU-X1 (10.000rpm):<br>dichromatic mirrors:<br>[http://www.semrock.com/FilterDetails.aspx?id=Di01-T405/488/568/647-13x15x0.5 Quad band T-405/488/568/647]<br> [http://www.semrock.com/FilterDetails.aspx?id=Di01-T405/488/561-13x15x0.5 Triple band T-405/488/561] <br>physical pinhole radius: 24um, physical pinhole spacing: 240um; back-projected pinhole radius: 0.15um, back-projected pinhole spacing: 1.5um (with 100x objective, 1.6x optovar)<br>Andor iXon EM+ DU-897 BV back illuminated EMCCD; pixel size of EMCCD chip: 16um<br>image pixel size (measured again after major scope repair on March 7th 2011): with 100x objective: 0.175um (was 0.178um before 20110307); 100x obj. and 1.6x optovar: 0.109um (was 0.111um before 20110307); 60x obj.: 0.289um (was 0.288um before); 60x obj. and 1.6x opt.: 0.180um (was 0.181um before 20110307); 10x obj. : 1.747um, 10x obj. and 1.6x opt.: 1.089um<br>camera projecting lens has about 0.92x (was 0.9x before 2011_03_07) magnification calculated from the measured pixel sizes
 
|reflectors=Optical Insight DualView image splitter (currently not installed): <br>holder for GFP/mCherry: BL 525/40; BS 565; ET 605/70<br>holder for GFP/RFP: HC 520/35; BS 565; HC 628/40<br>Sutter filter wheel: <br>position 0: 624/40,  1: open,  2: 525/30,  3: razor edge 568LP,  4: dual emitter: 512/23+630/91,  5: razor edge 488LP,  6: old CSU dual emitter:  525/30 + 650/?,  7: 605/70,  8: 445/40,  9: 716/40<br> microscope filter turret:<br> position 1: GFP cube,  2: RFP cube,  3: dual GFP + RFP cube,  4: cube with z375 RDC for micropoint cutter,  5: empty,  6: DIC analyzer
 
|reflectors=Optical Insight DualView image splitter (currently not installed): <br>holder for GFP/mCherry: BL 525/40; BS 565; ET 605/70<br>holder for GFP/RFP: HC 520/35; BS 565; HC 628/40<br>Sutter filter wheel: <br>position 0: 624/40,  1: open,  2: 525/30,  3: razor edge 568LP,  4: dual emitter: 512/23+630/91,  5: razor edge 488LP,  6: old CSU dual emitter:  525/30 + 650/?,  7: 605/70,  8: 445/40,  9: 716/40<br> microscope filter turret:<br> position 1: GFP cube,  2: RFP cube,  3: dual GFP + RFP cube,  4: cube with z375 RDC for micropoint cutter,  5: empty,  6: DIC analyzer
|features=-
+
|features=micro point with 15Hz 365nm cutter laser
 
|software=iQ 2.5.1
 
|software=iQ 2.5.1
 
|incubation=Stage incubator (T) + objective heater
 
|incubation=Stage incubator (T) + objective heater

Revision as of 16:24, 5 April 2012

Suitable for confocal imaging of EGFP, mCherry, dTomato, Draq5 and the like with high temporal and spatial resolution..
MPI19.jpg


Directions

[[image:|150px|text-bottom|right]]

MPI-CBG LMF.jpgMPI-CBG LMF (MPI-CBG, 038B)

Booking

https://python-srv1.mpi-cbg.de/lmf-ipf/cgi-bin/index.py

Details

microscope Olympus - IX71, Inverted stand, manual, 1.6x OptoVar, fast piezo objective z-positioner (Physik Instrumente - PI), Prior ProScanIII xy scanning stage
objectives
Olympus UPlanSApo 10x 0.4
Olympus UPlanApo 20x 0.7
Olympus UPlan FluarN 60x 0.9
Olympus UPlanSApo 60x 1.20 W
Olympus UPlanSApo 100x 1.4 Oil
illumination
Transmitted Light (Halogen)
Fluorescence (HBO)
Laser DPSS 488 nm
Laser DPSS 561 nm
Laser DPSS 640 nm
detection [[detection::Spinning disc scan head Yokogawa CSU-X1 (10.000rpm):
dichromatic mirrors:
Quad band T-405/488/568/647
Triple band T-405/488/561
physical pinhole radius: 24um, physical pinhole spacing: 240um; back-projected pinhole radius: 0.15um, back-projected pinhole spacing: 1.5um (with 100x objective, 1.6x optovar)
Andor iXon EM+ DU-897 BV back illuminated EMCCD; pixel size of EMCCD chip: 16um
image pixel size (measured again after major scope repair on March 7th 2011): with 100x objective: 0.175um (was 0.178um before 20110307); 100x obj. and 1.6x optovar: 0.109um (was 0.111um before 20110307); 60x obj.: 0.289um (was 0.288um before); 60x obj. and 1.6x opt.: 0.180um (was 0.181um before 20110307); 10x obj. : 1.747um, 10x obj. and 1.6x opt.: 1.089um
camera projecting lens has about 0.92x (was 0.9x before 2011_03_07) magnification calculated from the measured pixel sizes]]
reflectors Optical Insight DualView image splitter (currently not installed):
holder for GFP/mCherry: BL 525/40; BS 565; ET 605/70
holder for GFP/RFP: HC 520/35; BS 565; HC 628/40
Sutter filter wheel:
position 0: 624/40, 1: open, 2: 525/30, 3: razor edge 568LP, 4: dual emitter: 512/23+630/91, 5: razor edge 488LP, 6: old CSU dual emitter: 525/30 + 650/?, 7: 605/70, 8: 445/40, 9: 716/40
microscope filter turret:
position 1: GFP cube, 2: RFP cube, 3: dual GFP + RFP cube, 4: cube with z375 RDC for micropoint cutter, 5: empty, 6: DIC analyzer
features micro point with 15Hz 365nm cutter laser
software iQ 2.5.1
incubation Stage incubator (T) + objective heater
links
  • [- beam path]
  • [ advanced system check]
inv.nr. 0 10.1.65.208