SD1 - Andor Spinning Disc

From BioDIP
(Difference between revisions)
Jump to: navigation, search
Line 14: Line 14:
 
| objectives          = 10x / 0.30  Plan-Neofluar<br>60x / 1.20 W  UPlanSApo<br>100x / 1.40 Oil  UPlanSApo
 
| objectives          = 10x / 0.30  Plan-Neofluar<br>60x / 1.20 W  UPlanSApo<br>100x / 1.40 Oil  UPlanSApo
 
| illumination        = Halogen<br>HBO<br>488 nm solid-state laser Coherent Sapphire<br>561 nm solid-state laser Cobolt Jive
 
| illumination        = Halogen<br>HBO<br>488 nm solid-state laser Coherent Sapphire<br>561 nm solid-state laser Cobolt Jive
| detection            = Spinning disc scan head Yokogawa CSU10; back-projected pinhole size: 0.4um, back-projected pinhole spacing: 1.5um (with 100x objective, 1.6x optovar)<br>Andor iXon EM+ DU-897 BV back illuminated EMCCD; pixel size of EMCCD chip: 16um<br>image pixel size: with 100x objective: 0.16um; 100x obj. and 1.6x optovar: 0.1um; 60x obj.: 0.267um; 60x obj. and 1.6x opt.: 0.167um  
+
| detection            = Spinning disc scan head Yokogawa CSU10: dichromatic mirror: [http://www.semrock.com/Catalog/Detail.aspx?FilterPartID=855&CategoryID=82 T-405/488/568/647]<br> pinhole radius: 24um, pinhole spaceing: 240um; back-projected pinhole radius: 0.15um, back-projected pinhole spacing: 1.5um (with 100x objective, 1.6x optovar)<br>Andor iXon EM+ DU-897 BV back illuminated EMCCD; pixel size of EMCCD chip: 16um<br>image pixel size: with 100x objective: 0.16um; 100x obj. and 1.6x optovar: 0.1um; 60x obj.: 0.267um; 60x obj. and 1.6x opt.: 0.167um  
 
| reflectors          = Optical Insight DualView image splitter: <br>HC 520/35; BS 565; HC 628/40<br>Sutter filter wheel: <br>position 0: 624/40,  1: open,  2: 525/30,  3: razor edge 568LP,  4: dual emitter: 512/23+630/91,  5: razor edge 488LP,  6: old CSU dual emitter:  525/30 + 650/?,  7: blocked,  8: blocked,  9: 716/40<br> microscope filter turret:<br> position 1: GFP cube,  2: RFP cube,  3: dual GFP + RFP cube,  4: DAPI cube,  5: empty,  6: DIC analyzer  
 
| reflectors          = Optical Insight DualView image splitter: <br>HC 520/35; BS 565; HC 628/40<br>Sutter filter wheel: <br>position 0: 624/40,  1: open,  2: 525/30,  3: razor edge 568LP,  4: dual emitter: 512/23+630/91,  5: razor edge 488LP,  6: old CSU dual emitter:  525/30 + 650/?,  7: blocked,  8: blocked,  9: 716/40<br> microscope filter turret:<br> position 1: GFP cube,  2: RFP cube,  3: dual GFP + RFP cube,  4: DAPI cube,  5: empty,  6: DIC analyzer  
 
| features            =  
 
| features            =  

Revision as of 16:02, 22 January 2009

Suitable for confocal imaging of EGFP and mCherry with high temporal and spatial resolution..
MPI19.jpg


Directions

[[image:|150px|text-bottom|right]]

[[image:|100px]] MPI-CBG LMF (MPI-CBG, 038B)

Booking

https://python-srv1.mpi-cbg.de/lmf-ipf/cgi-bin/index.py

Details

microscope , inverted, manual, 1.6x optovar stand, Olympus IX71, fast piezo z-drive for objective
objectives
illumination
detection [[detection::Spinning disc scan head Yokogawa CSU10: dichromatic mirror: T-405/488/568/647
pinhole radius: 24um, pinhole spaceing: 240um; back-projected pinhole radius: 0.15um, back-projected pinhole spacing: 1.5um (with 100x objective, 1.6x optovar)
Andor iXon EM+ DU-897 BV back illuminated EMCCD; pixel size of EMCCD chip: 16um
image pixel size: with 100x objective: 0.16um; 100x obj. and 1.6x optovar: 0.1um; 60x obj.: 0.267um; 60x obj. and 1.6x opt.: 0.167um]]
reflectors Optical Insight DualView image splitter:
HC 520/35; BS 565; HC 628/40
Sutter filter wheel:
position 0: 624/40, 1: open, 2: 525/30, 3: razor edge 568LP, 4: dual emitter: 512/23+630/91, 5: razor edge 488LP, 6: old CSU dual emitter: 525/30 + 650/?, 7: blocked, 8: blocked, 9: 716/40
microscope filter turret:
position 1: GFP cube, 2: RFP cube, 3: dual GFP + RFP cube, 4: DAPI cube, 5: empty, 6: DIC analyzer