SD2 - Andor Spinning Disc FRAPPA

From BioDIP
(Difference between revisions)
Jump to: navigation, search
Line 3: Line 3:
 
| system-name          = Andor Spinning Disc FRAPPA
 
| system-name          = Andor Spinning Disc FRAPPA
 
| category            = Real-Time Confocal
 
| category            = Real-Time Confocal
| stand                = inverted, motorized, 2x optovar
+
| stand                = inverted
 
| image                = MPI12.jpg
 
| image                = MPI12.jpg
 
| facility            = MPI-CBG LMF
 
| facility            = MPI-CBG LMF
Line 11: Line 11:
 
| description          = fast, "real-time" confocal imaging
 
| description          = fast, "real-time" confocal imaging
 
| applications        = confocal imaging of DAPI,EGFP, mCherry, Cy5 with high temporal and spatial resolution.
 
| applications        = confocal imaging of DAPI,EGFP, mCherry, Cy5 with high temporal and spatial resolution.
| microscope          = Olympus IX81 with ZDC (hardware autofocus), Prior NanoScanZ, Prior ProScanII xy scanning stage  
+
| microscope          = motorized Olympus IX81 with ZDC (hardware autofocus), 2x OptoVar, Prior NanoScanZ, Prior ProScanII xy scanning stage  
 
| objectives          = 10x / 0.40  UPlanSApo<br>60x / 1.20 W  UPlanSApo<br>60x / 1.35 Oil  UPlanSApo
 
| objectives          = 10x / 0.40  UPlanSApo<br>60x / 1.20 W  UPlanSApo<br>60x / 1.35 Oil  UPlanSApo
 
| illumination        = Cairn OptoLED lite<br>HBO<br>405nm diode laser, 50mW<br>488 nm solid-state laser Coherent Sapphire, 50mW<br>561 nm solid-state laser Cobolt Jive, 50mW<br>640nm solid-state laser, 40mW
 
| illumination        = Cairn OptoLED lite<br>HBO<br>405nm diode laser, 50mW<br>488 nm solid-state laser Coherent Sapphire, 50mW<br>561 nm solid-state laser Cobolt Jive, 50mW<br>640nm solid-state laser, 40mW

Revision as of 16:16, 10 August 2009

Suitable for confocal imaging of DAPI,EGFP, mCherry, Cy5 with high temporal and spatial resolution..
MPI12.jpg


Directions

[[image:|150px|text-bottom|right]]

[[image:|100px]] MPI-CBG LMF (MPI-CBG, 038B)

Booking

https://python-srv1.mpi-cbg.de/lmf-ipf/cgi-bin/index.py

Details

microscope , inverted stand, motorized Olympus IX81 with ZDC (hardware autofocus), 2x OptoVar, Prior NanoScanZ, Prior ProScanII xy scanning stage
objectives
illumination
detection Spinning disc scan head Yokogawa CSU-X1; back-projected pinhole size: ??um, back-projected pinhole spacing: ??um (with 100x objective, 2x optovar)
Andor iXon EM+ DU-897 BV back illuminated EMCCD; pixel size of EMCCD chip: 16um
note there are two additional lenses in the optical path: 1.7x before and 1.2x after the CSU
image pixel size (calculated): with 60x objective: 0.131um; 60x obj. and 2.0x optovar: 0.065um; 40x obj.: 0.196um; 40x obj. and 2.0x opt.: 0.098um
reflectors Optical Insight DualView image splitter:
HC 520/35; BS 595 dcxr; HC 628/40
Sutter filter wheel:
position 0: 440/521/607/700 Quadband, 1: razor edge 488LP, 2: blocked, 3: razor edge 568LP, 4: blocked, 5: blocked, 6: open, 7: polarizer for DIC, 8: 525/30, 9: blocked
microscope filter turret:
position 1: empty, 2: DIC, 3: tripple DAPI + GFP + RFP cube, 4: GFP, 5: RFP, 6: DAPI cube (X 345/10; D 380LP; M 445/40)
features Andor FRAPPA
405nm pulsed diode laser from PicoQuant"Andor FRAPPA<br />405nm pulsed diode laser from PicoQuant" cannot be used as a page name in this wiki.
software
incubation stage incubator
links
  • [ manual]
  • [LPM laser power values]
  • [- beam path]
  • [ advanced system check]
inv.nr.



[[Category:{{{type}}}]]






Personal tools