SD2 - Andor Spinning Disc FRAPPA

From BioDIP
(Difference between revisions)
Jump to: navigation, search
(update of values for camera projecting lenses in front of Clara and iXon, moved old value to end of section)
(moved one more old value for iXon projecting lens to end of detection section)
Line 25: Line 25:
 
|ill6=Laser DPSS 640 nm
 
|ill6=Laser DPSS 640 nm
 
|ill7=Laser Diode 405 nm (pulsed)
 
|ill7=Laser Diode 405 nm (pulsed)
|detection=Spinning disc scan head Yokogawa CSU-X1; pinhole size: 24um, pinhole spacing: 240um; back-projected pinhole size: 0.4um, back-projected pinhole spacing: 4um (with 60x objective, 1x optovar)<br>Andor Dual camera port with two cameras: <br>Andor iXon EM+ DU-897 BV back illuminated EMCCD; pixel size of EMCCD chip: 16um<br>Andor Clara cooled interline CCD; pixel size of chip: 6.45um<br>note there are two additional lenses in the optical path: 1.0x before the CSU and another one after the CSU: a 2.475x in front of iXon  (was 1.9x before addition of dual camera switch)) in front of iXon or a 1.285x in front of Clara<br>image pixel size for use with iXon (measured): with 100x: 0.067(was 0.077um before Apr 2011); with 60x objective: 0.111 (was 0.129 before Apr 2011) (was 0.14um); 60x obj. and 2.0x optovar: was 0.0645um (was 0.07um); 40x obj.: was 0.194um (was 0.21um); 40x obj. and 2.0x opt.: was 0.0967um (was 0.11um) <br> pixel size with Clara (measured July 15th 2011, all with optovar 1x): 100x objective: 0.05um/px, 60x obj.: 0.084um/px, 10x obj: 0.511um/px <br> old values: <br>projecting lens in front of iXon: was 2.39x before removal of DualView on Dec 21st 2011; was 2.067x before addition of the dual filter wheel on Apr 28th 2011 <br> projecting lens in front of Clara: was 1.28x before removal of DualView on Dec 21st 2011; was 1.2x before addition of the dual filter wheel on Apr 28th 2011  
+
|detection=Spinning disc scan head Yokogawa CSU-X1; pinhole size: 24um, pinhole spacing: 240um; back-projected pinhole size: 0.4um, back-projected pinhole spacing: 4um (with 60x objective, 1x optovar)<br>Andor Dual camera port with two cameras: <br>Andor iXon EM+ DU-897 BV back illuminated EMCCD; pixel size of EMCCD chip: 16um<br>Andor Clara cooled interline CCD; pixel size of chip: 6.45um<br>note there are two additional lenses in the optical path: 1.0x before the CSU and another one after the CSU: a 2.475x in front of iXon or a 1.285x in front of Clara<br>image pixel size for use with iXon (measured): with 100x: 0.067(was 0.077um before Apr 2011); with 60x objective: 0.111 (was 0.129 before Apr 2011) (was 0.14um); 60x obj. and 2.0x optovar: was 0.0645um (was 0.07um); 40x obj.: was 0.194um (was 0.21um); 40x obj. and 2.0x opt.: was 0.0967um (was 0.11um) <br> pixel size with Clara (measured July 15th 2011, all with optovar 1x): 100x objective: 0.05um/px, 60x obj.: 0.084um/px, 10x obj: 0.511um/px <br> old values: <br>projecting lens in front of iXon: was 2.39x before removal of DualView on Dec 21st 2011; was 2.067x before addition of the dual filter wheel on Apr 28th 2011, was 1.9x before addition of dual camera switch on June 30th 2010 <br> projecting lens in front of Clara: was 1.28x before removal of DualView on Dec 21st 2011; was 1.2x before addition of the dual filter wheel on Apr 28th 2011
 
|reflectors=Sutter filter wheel: <br>position 0: 440/521/607/700 Quadband,  1: razor edge 488LP,  2: blocked,  3: razor edge 568LP,  4: Edge Basic longpass 635 LP,  5: BrightLine 445/45,  6: open,  7: polarizer for DIC,  8: 525/30,  9: BrightLine HC 607/36<br> microscope filter turret:<br> position 1: empty,  2: DIC,  3: tripple DAPI + GFP + RFP cube,  4: GFP,  5: RFP,  6: DAPI cube (X 345/10; D 380LP; M 445/40)
 
|reflectors=Sutter filter wheel: <br>position 0: 440/521/607/700 Quadband,  1: razor edge 488LP,  2: blocked,  3: razor edge 568LP,  4: Edge Basic longpass 635 LP,  5: BrightLine 445/45,  6: open,  7: polarizer for DIC,  8: 525/30,  9: BrightLine HC 607/36<br> microscope filter turret:<br> position 1: empty,  2: DIC,  3: tripple DAPI + GFP + RFP cube,  4: GFP,  5: RFP,  6: DAPI cube (X 345/10; D 380LP; M 445/40)
 
|features=-
 
|features=-

Revision as of 17:15, 4 January 2012

Suitable for confocal imaging, photobleaching and -activation of DAPI,EGFP, mCherry, Cy5 with high temporal and spatial resolution.
MPI12.jpg


Directions

[[image:|150px|text-bottom|right]]

[[image:|100px]] MPI-CBG LMF (MPI-CBG, 038B)

Booking

https://python-srv1.mpi-cbg.de/lmf-ipf/cgi-bin/index.py

Details

microscope Olympus - IX81, Inverted stand, motorized with ZDC (hardware autofocus), 2x Optovar, Prior NanoScanZ, Prior ProScanIII xy scanning stage
objectives
Olympus UPlanSApo 10x 0.4
Olympus UPlanSApo 60x 1.20 W
Olympus UPlanSApo 60x 1.35 Oil
Olympus UPlanSApo 100x 1.4 Oil
Olympus UPlanFluar 20x 0.5
illumination
Transmitted Light (LED)
Fluorescence (HBO)
Laser Diode 405 nm
Laser DPSS 488 nm
Laser DPSS 561 nm
Laser DPSS 640 nm
Laser Diode 405 nm (pulsed)
detection Spinning disc scan head Yokogawa CSU-X1; pinhole size: 24um, pinhole spacing: 240um; back-projected pinhole size: 0.4um, back-projected pinhole spacing: 4um (with 60x objective, 1x optovar)
Andor Dual camera port with two cameras:
Andor iXon EM+ DU-897 BV back illuminated EMCCD; pixel size of EMCCD chip: 16um
Andor Clara cooled interline CCD; pixel size of chip: 6.45um
note there are two additional lenses in the optical path: 1.0x before the CSU and another one after the CSU: a 2.475x in front of iXon or a 1.285x in front of Clara
image pixel size for use with iXon (measured): with 100x: 0.067(was 0.077um before Apr 2011); with 60x objective: 0.111 (was 0.129 before Apr 2011) (was 0.14um); 60x obj. and 2.0x optovar: was 0.0645um (was 0.07um); 40x obj.: was 0.194um (was 0.21um); 40x obj. and 2.0x opt.: was 0.0967um (was 0.11um)
pixel size with Clara (measured July 15th 2011, all with optovar 1x): 100x objective: 0.05um/px, 60x obj.: 0.084um/px, 10x obj: 0.511um/px
old values:
projecting lens in front of iXon: was 2.39x before removal of DualView on Dec 21st 2011; was 2.067x before addition of the dual filter wheel on Apr 28th 2011, was 1.9x before addition of dual camera switch on June 30th 2010
projecting lens in front of Clara: was 1.28x before removal of DualView on Dec 21st 2011; was 1.2x before addition of the dual filter wheel on Apr 28th 2011
reflectors Sutter filter wheel:
position 0: 440/521/607/700 Quadband, 1: razor edge 488LP, 2: blocked, 3: razor edge 568LP, 4: Edge Basic longpass 635 LP, 5: BrightLine 445/45, 6: open, 7: polarizer for DIC, 8: 525/30, 9: BrightLine HC 607/36
microscope filter turret:
position 1: empty, 2: DIC, 3: tripple DAPI + GFP + RFP cube, 4: GFP, 5: RFP, 6: DAPI cube (X 345/10; D 380LP; M 445/40)
features -
software [[software::iQ 1.10.5 user guide + iQ 2.4 (with Python)]]
incubation Stage incubator (T) + objective heater
links
  • [- beam path]
  • [ advanced system check]
inv.nr. 0 10.1.145.20