SD2 - Andor Spinning Disc FRAPPA

From BioDIP
(Difference between revisions)
Jump to: navigation, search
Line 12: Line 12:
 
| applications        = confocal imaging of DAPI,EGFP, mCherry, Cy5 with high temporal and spatial resolution.
 
| applications        = confocal imaging of DAPI,EGFP, mCherry, Cy5 with high temporal and spatial resolution.
 
| microscope          = Olympus IX81 with ZDC (hardware autofocus), Prior NanoScanZ, Prior ProScanII xy scanning stage  
 
| microscope          = Olympus IX81 with ZDC (hardware autofocus), Prior NanoScanZ, Prior ProScanII xy scanning stage  
| objectives          = 10x / 0.40  UPlanSApo<br>60x / 1.20 W  UPlanSApo<br>100x / 1.35 Oil  UPlanSApo
+
| objectives          = 10x / 0.40  UPlanSApo<br>60x / 1.20 W  UPlanSApo<br>60x / 1.35 Oil  UPlanSApo
| illumination        = Cairn OptoLED lite<br>HBO<br>405nm diode laser<br>488 nm solid-state laser Coherent Sapphire<br>561 nm solid-state laser Cobolt Jive<br>640nm solid-state laser
+
| illumination        = Cairn OptoLED lite<br>HBO<br>405nm diode laser, 50mW<br>488 nm solid-state laser Coherent Sapphire, 50mW<br>561 nm solid-state laser Cobolt Jive, 50mW<br>640nm solid-state laser, 40mW
| detection            = Spinning disc scan head Yokogawa CSU-X1; back-projected pinhole size: ??um, back-projected pinhole spacing: ??um (with 100x objective, 2x optovar)<br>Andor iXon EM+ DU-897 BV back illuminated EMCCD<br>
+
| detection            = Spinning disc scan head Yokogawa CSU-X1; back-projected pinhole size: ??um, back-projected pinhole spacing: ??um (with 100x objective, 2x optovar)<br>Andor iXon EM+ DU-897 BV back illuminated EMCCD; pixel size of EMCCD chip: 16um<br>note there are two additional lenses in the optical path: 1.7x before and 1.2x after the CSU<br>image pixel size (calculated): with 60x objective: 0.131um; 60x obj. and 2.0x optovar: 0.065um; 40x obj.: 0.196um; 40x obj. and 2.0x opt.: 0.098um <br>
 
| reflectors          = Optical Insight DualView image splitter: <br>HC 520/35; BS 595 dcxr; HC 628/40<br>Sutter filter wheel: <br>position 0: 440/521/607/700 Quadband,  1: razor edge 488LP,  2: blocked,  3: razor edge 568LP,  4: blocked,  5: blocked,  6: open,  7: polarizer for DIC,  8: 525/30,  9: blocked<br> microscope filter turret:<br> position 1: empty,  2: DIC,  3: tripple DAPI + GFP + RFP cube,  4: GFP,  5: RFP,  6: DAPI cube (X 345/10; D 380LP; M 445/40)
 
| reflectors          = Optical Insight DualView image splitter: <br>HC 520/35; BS 595 dcxr; HC 628/40<br>Sutter filter wheel: <br>position 0: 440/521/607/700 Quadband,  1: razor edge 488LP,  2: blocked,  3: razor edge 568LP,  4: blocked,  5: blocked,  6: open,  7: polarizer for DIC,  8: 525/30,  9: blocked<br> microscope filter turret:<br> position 1: empty,  2: DIC,  3: tripple DAPI + GFP + RFP cube,  4: GFP,  5: RFP,  6: DAPI cube (X 345/10; D 380LP; M 445/40)
 
| features            = Andor FRAPPA<br>405nm pulsed diode laser from PicoQuant
 
| features            = Andor FRAPPA<br>405nm pulsed diode laser from PicoQuant

Revision as of 14:27, 7 May 2009

Suitable for confocal imaging of DAPI,EGFP, mCherry, Cy5 with high temporal and spatial resolution..
MPI12.jpg


Directions

[[image:|150px|text-bottom|right]]

[[image:|100px]] MPI-CBG LMF (MPI-CBG, 038B)

Booking

https://python-srv1.mpi-cbg.de/lmf-ipf/cgi-bin/index.py

Details

microscope , inverted, motorized, 2x optovar stand, Olympus IX81 with ZDC (hardware autofocus), Prior NanoScanZ, Prior ProScanII xy scanning stage
objectives
illumination
detection Spinning disc scan head Yokogawa CSU-X1; back-projected pinhole size: ??um, back-projected pinhole spacing: ??um (with 100x objective, 2x optovar)
Andor iXon EM+ DU-897 BV back illuminated EMCCD; pixel size of EMCCD chip: 16um
note there are two additional lenses in the optical path: 1.7x before and 1.2x after the CSU
image pixel size (calculated): with 60x objective: 0.131um; 60x obj. and 2.0x optovar: 0.065um; 40x obj.: 0.196um; 40x obj. and 2.0x opt.: 0.098um
reflectors Optical Insight DualView image splitter:
HC 520/35; BS 595 dcxr; HC 628/40
Sutter filter wheel:
position 0: 440/521/607/700 Quadband, 1: razor edge 488LP, 2: blocked, 3: razor edge 568LP, 4: blocked, 5: blocked, 6: open, 7: polarizer for DIC, 8: 525/30, 9: blocked
microscope filter turret:
position 1: empty, 2: DIC, 3: tripple DAPI + GFP + RFP cube, 4: GFP, 5: RFP, 6: DAPI cube (X 345/10; D 380LP; M 445/40)
features Andor FRAPPA
405nm pulsed diode laser from PicoQuant"Andor FRAPPA<br />405nm pulsed diode laser from PicoQuant" cannot be used as a page name in this wiki.
software
incubation stage incubator
links
  • [ manual]
  • [LPM laser power values]
  • [- beam path]
  • [ advanced system check]
inv.nr.



[[Category:{{{type}}}]]






Personal tools