SD2 - Andor Spinning Disc FRAPPA

From BioDIP
Revision as of 18:18, 23 September 2009 by Old user (Talk | contribs)

Jump to: navigation, search
Suitable for confocal imaging of DAPI,EGFP, mCherry, Cy5 with high temporal and spatial resolution.
MPI12.jpg


Directions

[[image:|150px|text-bottom|right]]

[[image:|100px]] MPI-CBG LMF (MPI-CBG, 038B)

Booking

https://python-srv1.mpi-cbg.de/lmf-ipf/cgi-bin/index.py

Details

microscope , Inverted stand, motorized Olympus IX81 with ZDC (hardware autofocus), 2x OptoVar, Prior NanoScanZ, Prior ProScanII xy scanning stage
objectives
Olympus UPlanSApo 10x 0.4
Olympus UPlanSApo 60x 1.20 W
Olympus UPlanSApo 60x 1.35 Oil
illumination
Transmitted Light (LED)
Fluorescence (HBO)
Laser Diode 405 nm
Laser DPSS 488 nm
Laser DPSS 561 nm
Laser DPSS 640 nm
Laser Diode 405 nm (pulsed)
detection Spinning disc scan head Yokogawa CSU-X1; back-projected pinhole size: ??um, back-projected pinhole spacing: ??um (with 100x objective, 2x optovar)
Andor iXon EM+ DU-897 BV back illuminated EMCCD; pixel size of EMCCD chip: 16um
note there are two additional lenses in the optical path: 1.7x before and 1.2x after the CSU
image pixel size (calculated): with 60x objective: 0.131um; 60x obj. and 2.0x optovar: 0.065um; 40x obj.: 0.196um; 40x obj. and 2.0x opt.: 0.098um
reflectors Optical Insight DualView image splitter:
HC 520/35; BS 595 dcxr; HC 628/40
Sutter filter wheel:
position 0: 440/521/607/700 Quadband, 1: razor edge 488LP, 2: blocked, 3: razor edge 568LP, 4: blocked, 5: blocked, 6: open, 7: polarizer for DIC, 8: 525/30, 9: blocked
microscope filter turret:
position 1: empty, 2: DIC, 3: tripple DAPI + GFP + RFP cube, 4: GFP, 5: RFP, 6: DAPI cube (X 345/10; D 380LP; M 445/40)
features Andor FRAPPA
405nm pulsed diode laser from PicoQuant"Andor FRAPPA<br />405nm pulsed diode laser from PicoQuant" cannot be used as a page name in this wiki.
software
incubation Stage incubator (T) + objective heater
links
  • [- beam path]
  • [ advanced system check]
inv.nr. 0



Personal tools