Zeiss Spinning Disc - DZNE SD1
From BioDIP
(Difference between revisions)
Line 21: | Line 21: | ||
|ill2=Fluorescence (Metal Halide, HXP, 120W) | |ill2=Fluorescence (Metal Halide, HXP, 120W) | ||
|ill3=Laser Diode 405 nm | |ill3=Laser Diode 405 nm | ||
− | |ill4=Laser | + | |ill4=Laser OPSL 488nm |
|ill5=Laser DPSS 561 nm | |ill5=Laser DPSS 561 nm | ||
− | |ill6=Laser | + | |ill6=Laser Diode 635 nm |
− | |detection=Spinning disc scan head Yokogawa CSU- | + | |detection=Spinning disc scan head Yokogawa CSU-X1M 5000 dual cam; pinhole size: 24um, pinhole spacing: 240um |
− | Zeiss AxiCam | + | Zeiss AxiCam HRm Rev.3: 1388x1040 Px, pixel size 6.45x6.45um, 12f/sec at full chip |
Roper evolve 515 EMCCD with manual EM filter holder | Roper evolve 515 EMCCD with manual EM filter holder | ||
− | |reflectors= | + | |reflectors=Emissionsfilterrad |
− | + | Dichroic Quattro QDM 405+488+568+647 | |
− | + | Emission filter: DBP 527/54 + 645/60 | |
− | + | BP 450/50 | |
− | + | BP 525/50 | |
− | | | + | BP 629/62 |
− | |inv= | + | BP 690/50 |
+ | |features=- | ||
+ | |software=AxioVision Rel. 4.8.2 | ||
+ | |incubation=Not available | ||
+ | |inv=0 | ||
}} | }} |
Revision as of 16:56, 12 March 2012
Directions
Booking
https://techpool.biotec.tu-dresden.de/schedule/schedule.php?scheduleid=sc14ffd3aa3e44c3
Details
microscope | Zeiss - Axio Observer.Z1, Inverted stand, motorized, motorized xy scanning stage, motorized z-drive | ||
objectives | |||
illumination | |||
detection | Spinning disc scan head Yokogawa CSU-X1M 5000 dual cam; pinhole size: 24um, pinhole spacing: 240um
Zeiss AxiCam HRm Rev.3: 1388x1040 Px, pixel size 6.45x6.45um, 12f/sec at full chip Roper evolve 515 EMCCD with manual EM filter holder | ||
reflectors | Emissionsfilterrad
Dichroic Quattro QDM 405+488+568+647 Emission filter: DBP 527/54 + 645/60 BP 450/50 BP 525/50 BP 629/62 BP 690/50 | ||
features | - | ||
software | AxioVision Rel. 4.8.2 | ||
incubation | Not available | ||
links |
| ||
inv.nr. | 0 |