ASC Field Illumination

From BioDIP
(Difference between revisions)
Jump to: navigation, search
(images)
Line 25: Line 25:
 
<gallery>
 
<gallery>
 
file:Field illumination zeiss mirror.jpg|Zeiss mirror slide
 
file:Field illumination zeiss mirror.jpg|Zeiss mirror slide
file:Field illumination japanese flag.jpg|"Japanese flag", reflected light image of mirror slide, Zeiss LSM 510, 10x/0.45, zoom 0.7, 594 nm laser, HiLo LUT in Fiji
+
file:Field illumination japanese flag.jpg|"Japanese flag", reflected light image of mirror slide, Zeiss LSM 510, 10x/0.45, zoom 0.7, 594 nm laser, "HiLo" LUT
file:Field illumination surface plot.jpg|Fiji: surface plot of illumination pattern
+
file:Field illumination surface plot.jpg|Surface plot of illumination pattern
file:Field illumination inverted surface plot.jpg|Fiji: surface plot of illumination pattern, inverted
+
file:Field illumination inverted surface plot.jpg|Surface plot of illumination pattern, inverted
 +
file:Plot of 10xillumination594hft-405-488-594CH1.jpg|Line Profile of illumination pattern
 
</gallery>
 
</gallery>
 
__NOTOC__
 
__NOTOC__

Revision as of 09:41, 1 October 2009

The two images acquired via Fluorescence and Brightfield Imaging already gave an impression of the field illumination. Homogeneous illumination is particularly important for quantitative analysis, i.e. the comparison of intensities in different regions of the field of view. Therefore the following test provides a more quantitative analysis of the field illumination. Reasons for the illumination not being in the center might be:

  • lasers not coupled well into the fiber
  • fiber not coupled well into the scan-head
  • objective misaligned/damaged
  • pinhole not centered

Due to that, it makes sense to not only test 10x objective + 488 nm laser, but also other combinations. Try to compare images with open vs. closed pinhole to check for the pinhole alignment.

requirements

  • mirror slide
    • alternatively: reflective surface (i.e. glass slide, gold-coated surface)
    • for testing field illumination with immersion objectives, one should consider the use of the Leica Z-resolution mirror

acquisition

  • set the laser of interest for illumination
  • set pinhole to 1 AU
  • set up beam path for reflection (i.e. 80/20 splitter, AOBS in reflection mode)
  • focus on reflective side of the mirror
  • try to obtain "japanese flag"-like image (use i.e. "Range Indicator" LUT)
    • for quantitative measurements, do not oversaturate the center

analysis

  • Is the illumination centered well?
  • How big are the differences in intensity between center and edges?
  • ways of visualization:
    • Surface Plot: Fiji > Analyze > Surface Plot...