ASC Fluorescence and Brightfield Imaging

From BioDIP
(Difference between revisions)
Jump to: navigation, search
(images)
 
Line 15: Line 15:
 
File:Conva10xok-1.jpg with bar.jpg|Convallaria, Zeiss LSM 510, 10x/0.45 objective, 488 nm illumination, transmitted light detection
 
File:Conva10xok-1.jpg with bar.jpg|Convallaria, Zeiss LSM 510, 10x/0.45 objective, 488 nm illumination, transmitted light detection
 
</gallery>
 
</gallery>
 +
[[category:ASC]]
 
__NOTOC__
 
__NOTOC__

Latest revision as of 09:50, 1 October 2009

First step of the system check. The purpose is to test the general performance of the system: Do the basic functions work fine? Does the acquisition software behave as it should? Can one set up Koehler illumination? Are the transmitted and the reflected beam-paths ok? Although this test does not represent high-resolution work, it might reveal general problems with the system.

[edit] requirements

  • basic sample with fluorescence and contrast in transmitted light (i.e. Zeiss/Leica Convallaria demo sample)

[edit] acquisition

  • set up the beam-path for fluorescence and transmitted light
  • use basic air objective (i.e. 10x/0.4)
  • acquire two images in basic resolution (i.e. 512x512)

[edit] analysis

  • Does the illumination look reasonably homogeneous over the whole field of view in both images?
  • Is the transmitted light image free of dirt particles?
  • Are both images free of artifacts?

[edit] images