Chromatic aberration measurement and correction

From BioDIP
(Difference between revisions)
Jump to: navigation, search
(Created page with '*overlay UV/V and VIS (confocal systems with separate UV/V laser coupling) **purpose: check the coupling precision of the UV/V laser **requirements: 0.2/0.5µm fluorescent beads ...')
 
Line 3: Line 3:
 
**requirements: 0.2/0.5µm fluorescent beads sample
 
**requirements: 0.2/0.5µm fluorescent beads sample
 
**work flow: open pinhole(s), use high resolution apochromatic lens (NA 1.2 or above), acquire two images (UV/V plus VIS channel) with good sampling (pixel size ~100nm)
 
**work flow: open pinhole(s), use high resolution apochromatic lens (NA 1.2 or above), acquire two images (UV/V plus VIS channel) with good sampling (pixel size ~100nm)
 +
__NOTOC__

Revision as of 14:19, 30 September 2009

  • overlay UV/V and VIS (confocal systems with separate UV/V laser coupling)
    • purpose: check the coupling precision of the UV/V laser
    • requirements: 0.2/0.5µm fluorescent beads sample
    • work flow: open pinhole(s), use high resolution apochromatic lens (NA 1.2 or above), acquire two images (UV/V plus VIS channel) with good sampling (pixel size ~100nm)
Personal tools