MPI-CL1 - Leica TCS SP2 - startup & shutdown
From BioDIP
(Difference between revisions)
m |
m |
||
Line 2: | Line 2: | ||
=== control panel === | === control panel === | ||
− | [[image:mpi03_panel.jpg| | + | [[image:mpi03_panel.jpg|200x100px]] |
# Switch buttons from right to left (PC-Scanner-Laser) | # Switch buttons from right to left (PC-Scanner-Laser) | ||
# Before switching on the argon laser, check that the power Level is set to MIN. Once the laser is switched on you, can turn the power up (labeled mark). | # Before switching on the argon laser, check that the power Level is set to MIN. Once the laser is switched on you, can turn the power up (labeled mark). | ||
Line 8: | Line 8: | ||
# Switch on 543nm and 633nm laser line if needed. | # Switch on 543nm and 633nm laser line if needed. | ||
=== software initiation === | === software initiation === | ||
− | [[image:mpi03_setlimit.jpg| | + | [[image:mpi03_setlimit.jpg|200x100px]] [[image:mpi03_icon.jpg|200x100px]] [[image:mpi03_software.jpg|200x100px]] |
# Make sure that the upper stage limit is set before starting the Leica Confocal Software. | # Make sure that the upper stage limit is set before starting the Leica Confocal Software. | ||
# Choose “Personal” and start the initialization process. | # Choose “Personal” and start the initialization process. | ||
=== basic acquisition setup === | === basic acquisition setup === | ||
− | [[image:mpi03_swtoolbar.jpg| | + | [[image:mpi03_swtoolbar.jpg|200x100px]] [[image:mpi03_beampath.jpg|200x100px]] [[image:mpi03_controlpanel.jpg|200x100px]] |
# In the Aquire toolbar, select beam to open the Beam Path Setting window. | # In the Aquire toolbar, select beam to open the Beam Path Setting window. | ||
# There you can set up all the aquisition parameters according to your sample and needs. | # There you can set up all the aquisition parameters according to your sample and needs. | ||
Line 18: | Line 18: | ||
=== NOTE : This is a rather manual set up! YOU HAVE TO ... === | === NOTE : This is a rather manual set up! YOU HAVE TO ... === | ||
− | [[image:mpi03_stand.jpg| | + | [[image:mpi03_stand.jpg|200x100px]] |
* switch between scanning and visible mode ... | * switch between scanning and visible mode ... | ||
* change filter wheels (GFP, RFP filter for observation by eye, empty position for scanning) ... | * change filter wheels (GFP, RFP filter for observation by eye, empty position for scanning) ... | ||
* change objectives ... | * change objectives ... | ||
... BY HAND!!! | ... BY HAND!!! | ||
+ | |||
+ | [[category:hardware]] |
Revision as of 16:40, 19 January 2009
Contents |
How to initiate the Leica TCS SP2
control panel
- Switch buttons from right to left (PC-Scanner-Laser)
- Before switching on the argon laser, check that the power Level is set to MIN. Once the laser is switched on you, can turn the power up (labeled mark).
- Argon laser for laser lines: 458, 477, 488, 514. To switch on first turn the key to START, it then jumps automatically back to the ON position.
- Switch on 543nm and 633nm laser line if needed.
software initiation
- Make sure that the upper stage limit is set before starting the Leica Confocal Software.
- Choose “Personal” and start the initialization process.
basic acquisition setup
- In the Aquire toolbar, select beam to open the Beam Path Setting window.
- There you can set up all the aquisition parameters according to your sample and needs.
- Use the control panel to adjust for image gain and offset
NOTE : This is a rather manual set up! YOU HAVE TO ...
- switch between scanning and visible mode ...
- change filter wheels (GFP, RFP filter for observation by eye, empty position for scanning) ...
- change objectives ...
... BY HAND!!!