MPI-CL1 - Leica TCS SP2 - startup & shutdown

From BioDIP
(Difference between revisions)
Jump to: navigation, search
m
Line 1: Line 1:
 
== How to start the Leica TCS SP2 ==
 
== How to start the Leica TCS SP2 ==
  
=== control panel ===
+
=== Control Panel ===
 
[[image:mpi03_panel.jpg|200x100px]]
 
[[image:mpi03_panel.jpg|200x100px]]
 
# Switch buttons from right to left (PC-Scanner-Laser)  
 
# Switch buttons from right to left (PC-Scanner-Laser)  
Line 7: Line 7:
 
# Argon laser for laser lines: 458, 477, 488, 514. To switch on first turn the key to START, it then jumps automatically back to the ON position.
 
# Argon laser for laser lines: 458, 477, 488, 514. To switch on first turn the key to START, it then jumps automatically back to the ON position.
 
# Switch on 543nm and 633nm laser line if needed.
 
# Switch on 543nm and 633nm laser line if needed.
=== software initiation ===
+
=== Software Initiation ===
 
[[image:mpi03_setlimit.jpg|200x100px]] [[image:mpi03_icon.jpg|200x100px]] [[image:mpi03_software.jpg|200x100px]]
 
[[image:mpi03_setlimit.jpg|200x100px]] [[image:mpi03_icon.jpg|200x100px]] [[image:mpi03_software.jpg|200x100px]]
 
# Make sure that the upper stage limit is set before starting the Leica Confocal Software.
 
# Make sure that the upper stage limit is set before starting the Leica Confocal Software.
 
# Choose “Personal” and start the initialization process.
 
# Choose “Personal” and start the initialization process.
=== basic acquisition setup ===
+
=== Basic Acquisition Setup ===
 
[[image:mpi03_swtoolbar.jpg|200x100px]] [[image:mpi03_beampath.jpg|200x100px]] [[image:mpi03_controlpanel.jpg|200x100px]]
 
[[image:mpi03_swtoolbar.jpg|200x100px]] [[image:mpi03_beampath.jpg|200x100px]] [[image:mpi03_controlpanel.jpg|200x100px]]
 
# In the Aquire toolbar, select beam to open the Beam Path Setting window.  
 
# In the Aquire toolbar, select beam to open the Beam Path Setting window.  
Line 17: Line 17:
 
# Use the control panel to adjust for image gain and offset
 
# Use the control panel to adjust for image gain and offset
  
=== manual setup - hints ===
+
=== Manual Setup - Hints ===
 
[[image:mpi03_stand.jpg|200x100px]]
 
[[image:mpi03_stand.jpg|200x100px]]
  

Revision as of 17:19, 17 June 2009

Contents

How to start the Leica TCS SP2

Control Panel

Mpi03 panel.jpg

  1. Switch buttons from right to left (PC-Scanner-Laser)
  2. Before switching on the argon laser, check that the power Level is set to MIN. Once the laser is switched on you, can turn the power up (labeled mark).
  3. Argon laser for laser lines: 458, 477, 488, 514. To switch on first turn the key to START, it then jumps automatically back to the ON position.
  4. Switch on 543nm and 633nm laser line if needed.

Software Initiation

Mpi03 setlimit.jpg Mpi03 icon.jpg Mpi03 software.jpg

  1. Make sure that the upper stage limit is set before starting the Leica Confocal Software.
  2. Choose “Personal” and start the initialization process.

Basic Acquisition Setup

Mpi03 swtoolbar.jpg Mpi03 beampath.jpg Mpi03 controlpanel.jpg

  1. In the Aquire toolbar, select beam to open the Beam Path Setting window.
  2. There you can set up all the aquisition parameters according to your sample and needs.
  3. Use the control panel to adjust for image gain and offset

Manual Setup - Hints

Mpi03 stand.jpg


NOTE : This is a rather manual set up! YOU HAVE TO ...

  • switch between scanning and visible mode ...
  • change filter wheels (GFP, RFP filter for observation by eye, empty position for scanning) ...
  • change objectives ...

... BY HAND!!!