MPI-CO1 - Olympus FV1000 - startup & shutdown

From BioDIP
Revision as of 14:59, 6 February 2009 by Old user (Talk | contribs)

(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search

Contents

How to initiate the Olympus FV1000

hardware

  1. Switch on lasers you need:
    • 633nm laser: turn key
    • Argon laser: switch button, then turn the key
    • 561nm laser: press the green button
  2. Switch on the motorised stage
  3. Switch on the rest of the system using the power manager below the monitors. Think about what components you really need (otherwise leave them switched off):
    • LD405 ... 405nm imaging laser
    • 405nm cutter ... bleaching/ ablation laser

software

  1. Start software and log in as lmf user.
  2. If applicable, use the Olympus Dye list to help you set up the beam path settings.
  3. Set up the aquisition parameters according to your sample and needs. Once you have an image, use the HV and offset function to adjust for saturation and background.
  4. You might want to reuse parameters from an old image using the Explorer. Look for your image data, right mouse click on it and choose “Reload aquisition parameters”

basic troubleshooting

  • PROBLEM: No light at the object plane!
    • CHECK: excitation shutter of turret open??
  • PROBLEM: No 405nm imaging laser at the object plane!
    • CHECK: SIM function disabled??


software initiation

Mpi03 setlimit.jpg Mpi03 icon.jpg Mpi03 software.jpg

  1. Make sure that the upper stage limit is set before starting the Leica Confocal Software.
  2. Choose “Personal” and start the initialization process.