MPI-SD3 - Zeiss Spinning Disc - startup & shutdown
Contents |
Switching on the system
At first you have to switch on the ventilator. You can find the button on the right-hand side; push the button and regulate the airflow with the wheel. It is not necessary that you use full power - one-third is enough.
Every component you have to switch on is labeled with a number (from 1 to 8).
First, switch on the Argon laser (number 1) by pushing the black button to "on". Wait at least one minute, then turn the key around to "1". On the right you find the button for the laser module, switch it on (number 2).
Afterwards, switch on the three buttons (number 3,4 and 5) on the shelve above the monitor. Number 3 is the PC and camera, number 4 the definite focus and number 5 the metal halide lamp.
If you really need the metal halide lamp, switch it on separately.
Go on with the stand power supply 232 (number 6).
Think about which lasers you really need and switch them on (number 7). Please leave the key in the horizontal position, use only the switches.
Last, start the PC (number 8) and login.
After you have switched on all components you need to turn the trigger of the laser regulation unit to "laser run" in top position. You can do this shortly before you start your experiment. With the wheel you can regulate the power of the laser. It is strongly recommend that the control light is "green". Red light shows that you are using high power and this is going to age the laser very fast.
NOTE: After you have turned the key of the Argon laser to "on" you have to wait at least 5 minutes before turning the laser regulation module to "laser run".
If you need the CO2 and temperature module you can switch it on and off at every point of time. You can find it on the shelf behind the ventilator.
Switching off the system
Before you do anything, please check the LMF booking database at the END(!) of your imaging session to see whether you are the last user or not!!!
If you are not the last user...
- Close the main software window.
- Upload your data to your file server.
- Log off the windows account via "Start" > "Log off...".
If you are the last user...
- Close the main software window.
- Upload your data to your file server.
- Shut down windows via "Start" > "Shut down...".
- Switch off every component, starting with number 8 and ending with number 1 (backwards to the beginning).
NOTE: After the key of the "Lasos" box is turned to 0 wait at least 5 minutes before you are switching the laser unit off (black button) to allow the argon laser to cool down.
Remember to turn the laser regulation module back to standby and the wheel to minimum. Don't forget to switch off the ventilator and the metal halide lamp if you have used it