SD4 - Andor Revolution WD Borealis Mosaic

From BioDIP
(Difference between revisions)
Jump to: navigation, search
(change the iQ version to the current one)
 
(12 intermediate revisions by one user not shown)
Line 1: Line 1:
 
{{Equipment
 
{{Equipment
|system-number=MPI-SD4
+
|system-number=MPI- SD4
 
|system-name=SD4 Andor Revolution WD Borealis Mosaic
 
|system-name=SD4 Andor Revolution WD Borealis Mosaic
 
|type=Microscope
 
|type=Microscope
Line 11: Line 11:
 
|description=Fast, "real-time" spinning disc confocal imaging with photo-activation capability. Fast imaging of large field of view and large tile scans with even illumination.
 
|description=Fast, "real-time" spinning disc confocal imaging with photo-activation capability. Fast imaging of large field of view and large tile scans with even illumination.
 
|applications=confocal imaging, photo-activation using 405 laser with high temporal and spatial resolution; fast imaging of large samples over a long period of time
 
|applications=confocal imaging, photo-activation using 405 laser with high temporal and spatial resolution; fast imaging of large samples over a long period of time
|image=SD4.jpg
+
|image=SD4 system.jpg
 
|stand-name=Olympus - IX83
 
|stand-name=Olympus - IX83
 
|stand=Inverted
 
|stand=Inverted
Line 22: Line 22:
 
|obj6=Olympus UAPON 150x 1.45 TIRFM
 
|obj6=Olympus UAPON 150x 1.45 TIRFM
 
|ill1=Transmitted Light (LED)
 
|ill1=Transmitted Light (LED)
 +
|ill2=Fluorescence (LED)
 
|ill3=Laser Diode 405 nm
 
|ill3=Laser Diode 405 nm
 +
|ill4=Laser Diode 445 nm
 
|ill5=Laser DPSS 488 nm
 
|ill5=Laser DPSS 488 nm
 +
|ill6=Laser DPSS 514 nm
 
|ill7=Laser DPSS 561 nm
 
|ill7=Laser DPSS 561 nm
|ill8=Laser DPSS 639nm
+
|ill8=Laser DPSS 638nm
 +
|ill9=Laser DPSS 726nm
 
|detection=Spinning disc scan head Yokogawa CSU-W1 (4000rpm) with two pinhole discs (50um and 25um) and
 
|detection=Spinning disc scan head Yokogawa CSU-W1 (4000rpm) with two pinhole discs (50um and 25um) and
 
simultaneous dual camera detection: Camera 1 (Master): Andor iXon 888 Ultra with fringe suppression; Camera 2 (Slave): Andor iXon 888 Ultra
 
simultaneous dual camera detection: Camera 1 (Master): Andor iXon 888 Ultra with fringe suppression; Camera 2 (Slave): Andor iXon 888 Ultra
|software=Andor iQ version 3.3.1; MetaMorph version 7.8
+
|reflectors=CSU-W scan head:<br>
 +
3 main DM: Tripple (for 445nm, 515nm, 640nm laser); Quad (405nm, 488nm, 561nm, 640nm);
 +
Penta (405nm, 488nm, 561nm, 640nm, 720nm)<br>
 +
FW1 (Master): Pos(1) LP568, Pos(2) 525/50, Pos(3) 617/73, Pos(4) 607/36, Pos(5) 685/40, Pos(6) 785/62, Pos(7) pol, Pos(8) LP514, Pos(9) 542/27, Pos(10) 452/45 <br>
 +
FW1 (Slave): Pos(1) 465/30, Pos(2) 541/30, Pos(3) 525/50, Pos(4) 479/40, Pos(5) pol, Pos(6) - Pos(10) empty<br>
 +
 
 +
reflectors in IX83 stand:<br>
 +
DAPI set (Ex 377/50, DM 409, Em 447/60); GFP set (Ex 472/30, DM 495, Em 520/35); RFP set (Ex 562/40, DM 593, Em 624/40); CY5 set (Ex 628/40, DM 660, Em 692/40); Analyzer; DM for use with Mosaic 
 +
|software=Andor iQ version 3.6; MetaMorph version 7.8<br>
 +
communication connections:<br>COM1(Port1): IX83; COM3: CSU-WD; COM4: Mosaic shutter; COM6: prior XY stage; COM7: 16 channel CooLED; <br>Device: Mosaic; Device1: Mosaic laser; DDA06/16 DAC board: Piezo, trigger box; ILE3082M-ILE3082S: ILE master, slave
 
|incubation=Cage incubator (T+CO2)
 
|incubation=Cage incubator (T+CO2)
 +
|link1=https://docs.google.com/spreadsheets/d/1UNHFIrXe5ojXuf_fcb3H5PjxFdP4PqSeP5-Yi8-7UaE/pubhtml
 
}}
 
}}

Latest revision as of 11:59, 8 June 2020

Suitable for confocal imaging, photo-activation using 405 laser with high temporal and spatial resolution; fast imaging of large samples over a long period of time.
SD4 system.jpg


[edit] Directions

Logo MPI-CBG.jpg
MPI-CBG-Lmf-logo-big.pngMPI-CBG LMF (MPI-CBG, 036)

[edit] Booking

https://python-srv1.mpi-cbg.de/lmf-ipf/cgi-bin/index.py

[edit] Details

microscope Olympus - IX83, Inverted stand, motorized with ZDC (hardware autofocus and focus search), Prior NanoScanZ (400um travel range), Prior ProScanIII xy scanning stage
objectives
Olympus UPlan SApochromat 10x 0.4
Olympus UPlanSApo 30x 1,05 S
Olympus UPLSAPO 40x 1.25 SIL
Olympus UPLSAPO 60x 1.3 SIL
Olympus UPlan SApochromat 100x 1.4 Oil
Olympus UAPON 150x 1.45 TIRFM
illumination
Transmitted Light (LED)
Fluorescence (LED)
Laser Diode 405 nm
Laser Diode 445 nm
Laser DPSS 488 nm
Laser DPSS 514 nm
Laser DPSS 561 nm
Laser DPSS 638nm
Laser DPSS 726nm
detection Spinning disc scan head Yokogawa CSU-W1 (4000rpm) with two pinhole discs (50um and 25um) and

simultaneous dual camera detection: Camera 1 (Master): Andor iXon 888 Ultra with fringe suppression; Camera 2 (Slave): Andor iXon 888 Ultra

reflectors CSU-W scan head:

3 main DM: Tripple (for 445nm, 515nm, 640nm laser); Quad (405nm, 488nm, 561nm, 640nm); Penta (405nm, 488nm, 561nm, 640nm, 720nm)
FW1 (Master): Pos(1) LP568, Pos(2) 525/50, Pos(3) 617/73, Pos(4) 607/36, Pos(5) 685/40, Pos(6) 785/62, Pos(7) pol, Pos(8) LP514, Pos(9) 542/27, Pos(10) 452/45
FW1 (Slave): Pos(1) 465/30, Pos(2) 541/30, Pos(3) 525/50, Pos(4) 479/40, Pos(5) pol, Pos(6) - Pos(10) empty

reflectors in IX83 stand:
DAPI set (Ex 377/50, DM 409, Em 447/60); GFP set (Ex 472/30, DM 495, Em 520/35); RFP set (Ex 562/40, DM 593, Em 624/40); CY5 set (Ex 628/40, DM 660, Em 692/40); Analyzer; DM for use with Mosaic

features
software Andor iQ version 3.6; MetaMorph version 7.8

communication connections:
COM1(Port1): IX83; COM3: CSU-WD; COM4: Mosaic shutter; COM6: prior XY stage; COM7: 16 channel CooLED;
Device: Mosaic; Device1: Mosaic laser; DDA06/16 DAC board: Piezo, trigger box; ILE3082M-ILE3082S: ILE master, slave"Andor iQ version 3.6; MetaMorph version 7.8<br /> communication connections:<br />COM1(Port1): IX83; COM3: CSU-WD; COM4: Mosaic shutter; COM6: prior XY stage; COM7: 16 channel CooLED; <br />Device: Mosaic; Device1: Mosaic laser; DDA06/16 DAC board: Piezo, trigger box; ILE3082M-ILE3082S: ILE master, slave" cannot be used as a page name in this wiki.

incubation Cage incubator (T+CO2)
links
  • [ beam path]
  • [ advanced system check]
inv.nr.