Suitable for multi-channel imaging of fixed sample, sample screening, including CFP/Yfp FRET, image acquisition for deconvolution and DIC imaging.
MPI07.jpg


Directions

[[image:|150px|text-bottom|right]]

[[image:|100px]] MPI-CBG LMF (MPI-CBG, U48)

Booking

https://python-srv1.mpi-cbg.de/lmf-ipf/cgi-bin/index.py

Details

microscope Zeiss - Axioplan 2, Upright stand, manual XY stage, motorized Z-drive, fluorescence, transmitted light with manual DIC
objectives
Zeiss Plan-Neofluar 100x 1.3 Oil
illumination
Transmitted Light (Halogen)
Fluorescence (HBO)
detection Diagnostic Instruments RT SPOT monochrome camera with 1600x1200 pixels (6.7 um^2)
reflectors DAPI (X BP 333-383, D LP 408, M BP 432-484)
FITC (X BP 446-494, DLP 513, M BP 516-567)
TRITC ( X BP 543-554, D LP 578, M BP 579-637
YFP 491-509, D LP 517, M BP 519-546)
CFP X BP 426-446, D LP 453,M BP 465-499)
Brightfield, Darkfield, Phase Contrast, and DIC
FITC/TRITC (X DBP 474-490 & 542-566, D DBP 502-533 & 577-677< M DBP 503-533 & 581-621)
features -
software -
incubation Not available
links
  • [ manual]
  • [- laser power values]
  • [- beam path]
  • [ advanced system check]