how to to do a proper tile scan using multi-time module
 
how to to do a proper tile scan using multi-time module
  
 +
*set up imaging conditions
 +
*save scan configuration
 +
*generate database file for the tile scan
 +
*open multi-time > multiple locations
 +
*define configuration > load configuration
 +
*a) edit locations > tile mark corners > overlap 20% > create tile locations
 +
*b) edit locations > tile > define tile numbers > overlap 20% > create tile locations
 +
*start time
 +
*wait for acquisition and saving
 +
*select resulting window All_(filename)_Sum
 +
*tile utilities > correlate > correlation range 3 um
 +
*wait for correlation
 +
*save the result
  
 
[[Image:Zeiss_lsm510_filemenu.jpg|thumb]]
 
[[Image:Zeiss_lsm510_filemenu.jpg|thumb]]

Revision as of 16:15, 20 June 2008

manual

Zeiss LSM 510

how to to do a proper tile scan using multi-time module

  • set up imaging conditions
  • save scan configuration
  • generate database file for the tile scan
  • open multi-time > multiple locations
  • define configuration > load configuration
  • a) edit locations > tile mark corners > overlap 20% > create tile locations
  • b) edit locations > tile > define tile numbers > overlap 20% > create tile locations
  • start time
  • wait for acquisition and saving
  • select resulting window All_(filename)_Sum
  • tile utilities > correlate > correlation range 3 um
  • wait for correlation
  • save the result