MPI-CBG LMF
+
This section suggests protocols for the evaluation of high end light microscopes. The tests should be carried out in a regular manner. Although the protocols are set up for the use with [[confocal]] laser scanning microscopes (CLSM), most of them can be adopted for the use with wide-field systems and other types of light microscopes.
 
+
# [[ASC Fluorescence and Brightfield Imaging|Fluorescence and Brightfield Imaging]]
Advanced System Check
+
# [[ASC Field Illumination|Field Illumination]]
 
+
# [[ASC Color Overlay|Color Overlay]]
Protocol (draft)
+
# [[ASC Point Spread Function|Point Spread Function]]
 
+
# [[ASC Z-Resolution|Z-Resolution]]
*fluorescence and bright field imaging (all systems)
+
The individual protocols are optimized for PDF output via the "Print as PDF"-feature in the sidebar.
**purpose: test general performance of the system - basic functions work fine?
+
[[category:Manuals|A]]
**requirements: Zeiss/Leica Convallaria demo sample
+
[[category:ASC]]
**work flow: acquire two images in basic resolution (512x512), blue ex / green em plus transmitted light image, using 10x objective
+
 
+
*psf
+
**purpose:
+
**requirements:
+
**work flow:
+
*overlay UV/V and VIS (confocal systems with separate UV/V laser coupling)
+
**purpose: check the coupling precision of the UV/V laser
+
**requirements: 0.2/0.5µm fluorescent beads sample
+
**work flow: open pinhole(s), use high resolution apochromatic lens (NA 1.2 or above), acquire two images (UV/V plus VIS channel) with good sampling (pixel size ~100nm)
+

Latest revision as of 14:04, 1 October 2009

This section suggests protocols for the evaluation of high end light microscopes. The tests should be carried out in a regular manner. Although the protocols are set up for the use with confocal laser scanning microscopes (CLSM), most of them can be adopted for the use with wide-field systems and other types of light microscopes.

  1. Fluorescence and Brightfield Imaging
  2. Field Illumination
  3. Color Overlay
  4. Point Spread Function
  5. Z-Resolution

The individual protocols are optimized for PDF output via the "Print as PDF"-feature in the sidebar.