This section suggests protocols for the evaluation of high end light microscopes. The tests should be carried out in a regular manner. Although the protocols are set up for the use with [[confocal]] laser scanning microscopes (CLSM), most of them can be adopted for the use with wide-field systems and other types of light microscopes.
[[ASC Field Illumination]]
+
# [[ASC Fluorescence and Brightfield Imaging|Fluorescence and Brightfield Imaging]]
[[ASC Color Overlay]]
+
# [[ASC Field Illumination|Field Illumination]]
[[ASC Point Spread Function]]
+
# [[ASC Color Overlay|Color Overlay]]
[[ASC Z-Resolution]]
+
# [[ASC Point Spread Function|Point Spread Function]]
 +
# [[ASC Z-Resolution|Z-Resolution]]
 +
The individual protocols are optimized for PDF output via the "Print as PDF"-feature in the sidebar.
 +
[[category:Manuals|A]]
 +
[[category:ASC]]

Latest revision as of 14:04, 1 October 2009

This section suggests protocols for the evaluation of high end light microscopes. The tests should be carried out in a regular manner. Although the protocols are set up for the use with confocal laser scanning microscopes (CLSM), most of them can be adopted for the use with wide-field systems and other types of light microscopes.

  1. Fluorescence and Brightfield Imaging
  2. Field Illumination
  3. Color Overlay
  4. Point Spread Function
  5. Z-Resolution

The individual protocols are optimized for PDF output via the "Print as PDF"-feature in the sidebar.